Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8020B: CAL ADJ /5500 DATE: 16-Jan-95 AUTHOR: Fluke Corp. REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 73 CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # Fluke 8020B instruction manual (PN 616029 July 1981). # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 55_v8020.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5500A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- J F W 1.002 HEAD EQUIPMENT SETUP 1.003 HEAD CALIBRATION ACCESS PROCEDURE 1.004 DISP Disconnect the test leads and the battery eliminator 1.004 DISP if attached. Open the battery compartment, and 1.004 DISP disconnect the battery. Remove the three phillips-head 1.004 DISP screws from the bottom of the case. Turn the instrument 1.004 DISP face-up, and grasp the top cover at both sides of the 1.004 DISP input connectors. Then, pull the top cover from the UUT. 1.005 DISP Re-connect the battery. 1.005 DISP Ambient temperature: 21C - 25C. 1.005 DISP Relative humidity: less than 80%. 1.005 DISP Allow the UUT 30 min. to stabilize. 1.006 DISP Turn the UUT ON. 1.007 HEAD DC VOLTAGE CALIBRATION ADJUSTMENT 1.008 PIC 55_V8021 1.009 RNG 200mV 1.010 5500 190mV S 2W 1.011 DISP Adjust R5 for a reading of 190.0mV exactly. 1.012 5500 * S 1.013 HEAD AC VOLTAGE CALIBRATION ADJUSTMENT 1.014 RNG 200mV AC 1.015 5500 190mV 100H SI S 2W 1.016 DISP Adjust R9 for a reading of 190.0mV (+/- 1/2 count). 1.017 5500 * S 1.018 RNG 2V AC 1.019 5500 1.9V 5kH SI S 2W 1.020 DISP Adjust C1 for a reading of 1.900V (+/- 5 counts). 1.021 END